Standard Operation Procedure of Bacterial Endotoxin Test for medical devices:


Standard Operation Procedure of Bacterial Endotoxin Test for medical devices:

SCOPE

The purpose of this procedure is to provide a method for the routine testing of Medical devices for the presence of bacterial endotoxins.

MATERIALS & EQUIPMENT


1.                  Limulus Amebocyte Lysate (LAL) Reagent /Tachypleus Amebocyte Lysate    
            (TAL)  Reagent
2                    Control Standard Endotoxin (CSE) with a Certificate of Analysis stating 
            the  potency of the CSE in EU/ng for the specific lots of LAL & CSE
            reagents in use in the laboratory.
3.            LAL Reagent Water/BET water.
4.            Sterile & Endotoxin Free Pipettes.
5.            Endotoxin Free dilution tubes (borosilicate glass)
6.            Calibrated mechanical pipettors & sterile disposable tips
7.            Vortex mixer



PROCEDURE

Prerequisites


All testing shall be performed by qualified analysts, using lysate with a confirmed label claim sensitivity.
Glassware used in the performance of the test must be depyrogenated.

Sampling


Select not less than 3 devices and up to 10 devices. Extract it with suitable volume of rinsing solution like LRW or physiological saline and allow it to stand at room temperature for one hour.

Reagents


Lysate shall be reconstituted with LRW as per manufacturer’s instructions, and labeled with the date of reconstitution.
Lyophilized LAL may be stored at 2-8 deg.C. Once reconstituted, the lysate may be kept under refrigeration(2-8 deg.C) for up to 24 hours.
Reconstituted reagent may be frozen at –20 deg.C or below for up to 28 days. Reagent may be frozen and thawed only once.

Control Standard Endotoxin


CSE shall be reconstituted and vortexed as per manufacturer’s instructions

Once reconstituted, the stock solution of CSE may be kept under refrigeration
(2-8deg.C) for up to four weeks

Refer to the relevant CoA & reconstitute the CSE vial with specified volume of LRW.

Vortex this stock solution on a Vortex mixer for at least 5 min.

Arrange dilution tubes on a test tube rack and label them (2l, l, ½l & ¼ l) where l indicates the sensitivity of LAL Reagent.

Prepare serial dilutions of CSE in LRW in the dilution tubes marked 2l, l, ½l & ¼l.

Test for Sensitivity of the Lysate


The label claim sensitivity of the lysate is referred to as lambda.(l)

A serial two fold dilution series of CSE shall be prepared in LRW to bracket the lysate sensitivity.(2l,l,l/2, l/4)

The label claim sensitivity of the lysate is confirmed if the end point for the dilution series confirm lambda +/- one twofold dilution.

Formula and calculations

Endotoxin limit = K x N / V

Where K – amount of EU/device

            N --  No.of devices tested

            V – Total volume of the rinse

Maximum Valid Dilution (MVD)

MVD = Endotoxin limit / Lysate sensitivity(l)






Routine Testing


Preparation of Controls for testing

The label claim sensitivity of the lysate is referred to as lambda. A serial twofold dilution series of CSE shall be prepared in LRW to bracket the lysate sensitivity.

A 2l positive product control (PPC) for each lot of product under test shall be prepared.  The PPC acts as a control to check for inhibition of the LAL test by the product dilution under test.  The PPC shall be tested in duplicate.

A negative control (LRW) shall be tested in duplicate.

A 2l endotoxin / water control must be included with each test run that does not have its own concurrent control dilution series.

Test Procedure


100mL of eluate from a pooled sample (diluted to not more than the  MVD) is added to the respective reaction tube using a calibrated mechanical pipettor fitted with a disposable tip.

100mL of reconstituted LAL reagent is added to each tube in turn, starting with the negative control.

The tubes are then inserted into the heating block where they will incubate for a period of 60+/- 2 minutes.

After incubation, tubes are removed from the heating block one at a time, and smoothly rotated 180 deg, and scored (and documented) as positive or negative.

Interpretation


The preparation being examined complies with the test if negative results are found for both test replicates.

The preparation being examined does not comply with the test if positive results  are found in both test replicates.

Evaluation


The endotoxin content is to be reported in terms of EU/device using the formula
Endotoxin content in EU/device =
Endotoxin value (in EU/mL) x Volume of rinsing solution per device.                                                                                   
                                           ---------------------------